Description / Abstract:
This test method covers the detection of holes in medical
gloves.
This test method is limited to the detection of holes that allow
water leakage under the conditions of the test.
The smallest hole size that will allow water leakage in a
medical glove has not been determined and is beyond the scope of
this test method.
The safe and proper use of medical gloves is beyond the scope of
this test method.
The values stated in SI units are to be regarded as standard. No
other units of measurement are included in this standard.
This standard does not purport to address all of the safety
concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and
health practices and determine the applicability of regulatory
limitations prior to use.