This part of IEC 62366 specifies a PROCESS for a MANUFACTURER to
analyze, specify, develop and evaluate the USABILITY of a MEDICAL
DEVICE as it relates to SAFETY. This USABILITY ENGINEERING (HUMAN
FACTORS ENGINEERING) PROCESS permits the MANUFACTURER to assess and
mitigate RISKS associated with CORRECT USE and USE ERRORS, i.e.,
NORMAL USE. It can be used to identify but does not assess or
mitigate RISKS associated with ABNORMAL USE.
NOTE 1 SAFETY is freedom from unacceptable RISK. Unacceptable
RISK can arise from USE ERROR, which can lead to exposure to direct
physical HAZARDS or loss or degradation of clinical
functionality.
NOTE 2 Guidance on the application of USABILITY ENGINEERING to
MEDICAL DEVICES is available in IEC 62366-22, which addresses not
only SAFETY but also aspects of USABILITY not related to
SAFETY.
If the USABILITY ENGINEERING PROCESS detailed in this
International Standard has been complied with, then the USABILITY
of a MEDICAL DEVICE as it relates to SAFETY is presumed to be
acceptable, unless there is OBJECTIVE EVIDENCE to the contrary.
NOTE 3 Such OBJECTIVE EVIDENCE can subsequently originate from
POST-PRODUCTION surveillance.