This International Standard specifies a four-pole connector
system for implantable cardiac rhythm management devices which have
pacing, electrogram sensing and/or defibrillation functions. This
International Standard includes requirements for the connector
portion of an implantable lead as well as for the mating connector
cavity attached to an implantable pulse generator. Essential
dimensions and performance requirements are specified together with
appropriate test methods.
This International Standard is not intended to replace or
provide alternatives for unipolar or bipolar connector standards
that currently exist (such as ISO 11318 and ISO 5841-3). This
International Standard is not applicable to high voltage systems
with intended outputs greater than 1,000 V and/or 50 A. This
International Standard is not applicable to systems which include
sensors or unique electrodes that are not capable of conventional
pacing, electrogram sensing and/or defibrillation functions.
This International Standard does not specify all connector
features. It does not address all aspects of functional
compatibility, safety or reliability of leads and pulse generators
assembled into a system.
NOTE Lead and pulse generator connector systems not conforming
to this International Standard might be safe and reliable, and
might have clinical advantages.