Description / Abstract:
This International Standard specifies the requirements for test
equipment to be used to test chemical and biological indicators for
steam, ethylene oxide, dry heat and vaporized hydrogen peroxide
processes for conformity to the requirements given in ISO 11140-1
for chemical indicators, or the requirements given in the ISO 11138
series for biological indicators. This International Standard also
provides informative methods useful in characterizing the
performance of biological and chemical indicators for intended use
and for routine quality control testing.
ISO 11138-2, ISO 11138-3, ISO 11138-4, and ISO 11140-1 require
the use of resistometers specified in this International Standard,
and these resistometers are used in conjunction with the test
methods specified in the appropriate parts of ISO 11138 and ISO
11140.
NOTE Resistometers for formaldehyde indicators are not included
in this International Standard. Test methods using laboratory
apparatus for steam-formaldehyde are included in ISO 11138-5, ISO
11140-3, and ISO 11140-4.
This International Standard does not address the methods used to
demonstrate compliance of biological or chemical indicators to ISO
11138 and ISO 11140, as these are covered in the appropriate parts
of these standards. Indicators used with combination processes,
such as washer-disinfection, are not covered by this International
Standard.
NOTE Test equipment and methods necessary for ISO 11140-3, ISO
11140-4 or ISO 11140-5 are specified in those standards.
This International Standard does not address safety aspects of
the test equipment because these are usually covered by specific
regional, national or local regulations.